You are here: Home > News > Company news

The Search for Standard Sound Insulation Samples in Impedance Tubes

2026-03-19

Unlike sound pressure level measurement, there is no calibrator for measuring sound absorption coefficient and sound transmission loss. Impedance tube calibration is usually performed using standard specimens. It has been fifteen years since the launch of Soundwel impedance tubes. At that time, there was no national standard for sound insulation in China, nor were there recommended standard specimens for sound absorption. Referring to foreign products, we adopted a type of sponge with good sound absorption performance as the sound absorption standard specimen.


Soundwel 100mm Impedance Tube



Under varying factory conditions with different temperatures and humidity at different times, the sound absorption test results of the same material using 10 different test systems are shown in the figure above, with very high consistency. The error is smaller than that specified in JJF 1446-2014 Calibration Specification for Impedance Tubes (Transfer Function Method). After the release of GB/Z 27764-2014 Acoustics — Measurement of sound transmission loss in impedance tubes in 2011, sound transmission loss measurements using impedance tubes became more common.


First, we need to understand sound insulation of materials. For a homogeneous panel, the typical sound insulation curve is shown in the figure below. From low to high frequencies, there are four regions:stiffness-controlled region, damping-controlled region, mass-controlled region, and resonance-controlled region. The first transition frequency between the stiffness-controlled region and the damping-controlled region is the first-order mode of the structure.This means that the higher the first-order frequency of the structure, the larger the proportion occupied by the stiffness-controlled region.



Generally, we expect the sound transmission loss measured in the impedance tube to represent the actual sound insulation performance of the structure. The sound insulation in the stiffness‑controlled region at low frequencies is not meaningful.


Impedance tubes come in two sizes:a large tube (100mm in diameter) and a small tube (30mm in diameter). Sample dimensions are set according to the tube diameter, which means the samples are necessarily very small.

When a sample has very high stiffness, its boundary conditions and high stiffness result in an extremely high first‑order frequency.In this case, the measured sound insulation almost entirely falls in the stiffness‑controlled region and is not accurate. Simply put:reducing the sample size causes large discrepancies between the stiffness‑controlled and damping‑controlled regions measured in the tube,compared with the sound insulation of the same material used in large‑area applications.


Therefore, when selecting samples, we aim to make the first‑order frequency of the sample sufficiently low.We must choose samples with low stiffness and softness,so that the measured sound insulation curve can effectively avoid the stiffness‑controlled region.


Moreover, since US standards are more tolerant at low frequencies,standard sound insulation specimens only need to meet the requirements of the small tube in practice.(The small tube can measure down to 150 Hz, which is already beyond the stiffness‑controlled region.) Before this, we wanted to find a standard sound insulation sample, so we conducted tests.



However, the consistency was always poor whenever measurements were taken by inserting samples directly into the tube.Even the method of sandwiching a silicone sheet between two sponges imposed excessively strict requirements on sample cutting precision and the consistency of the two sponges.As a result, all attempts at measuring Q ended in failure.



As shown in the figure above, other manufacturers also install sound insulation materials by pushing them into the impedance tube, which leads to significant edge effects and poor measurement consistency. Poor sealing will result in even worse data.


The standard requires sealing the sample edges with grease or tape. However, as a standard sample, we do not want users to perform extra operations. Therefore, following the recommendations in Appendix C of the standard, we selected a soft, 1 mm thick, white, odorless silicone sheet as the standard sound insulation specimen. The specimen has a diameter larger than the tube inner diameter and is fixed by edge clamping during testing.






Photo of Specimen Installation

The sound insulation curve close to the theoretical value finally obtained is shown below.



The test results of different materials and different fillings are as follows.



It seems that this is the standard sound insulation test piece we are looking for. It is not easily affected by temperature and humidity, has no unpleasant smell, is soft and easy to install. Later, we will customize standard-sized mechanically processed samples for multiple tests and compare them with the test data of other pipes. According to the standards, we pay more attention to the part above the quality control zone above 300Hz.


In the future, when you purchase the impedance tubes of BSWA Technology will provide you with standard sound insulation and sound absorption samples for calibration.